TE-6070V High Volume PM-10 VFC Monitor in Ambient Air

     TE-6070V                                PM-10 High Volume Air Sampler: Complete With
                                                        Size selective inlet, outdoor anodize aluminum shelter,
                                                        aluminum blower motor assembly, 8" x 10" stainless steel
                                                        filter holder, volumetric flow controller with flow look-up
                                                        table, 7 day mechanical timer, elapsed time indicator,
                                                        continuous flow/pressure recorder, 30" water manometer and
                                                        filter paper cartridge(110v/60hz or 220v/50hz)

SPECIFICATIONS

  Inlet collection efficiency

  Meet EPA'S PM-10 Code of Federal Regulations

  Appendix J to part 50

  Flow rate

  40 cfm (adjustable) (1.02 to 1.24§©)

  Filter media

  Micro-Quartz filter media 8" * 10" for PM10 (TE-QMA)

  Flow control

  VFC (TE-10557)

  Water Manometer

  2-stage vacuum 1.0 HP 110v/60hz(TE-115923)

                                       220v/50hz(TE-116111)

  Flow Indicator

  Flow/pressure recorder (TE-5009) with TE-106 Charts

  Mechanical programmable timer

  7-day mechanical timer (TE-5007)

  Elapsed time indicator device

  Elapsed time indicator (TE-5012)

  Shipping Information

 

  Size

  22.5" * 20"* 45.5" sampler carton

  32"* 32" * 25.5" inlet carton

  Weight

  65 lbs-sampler carton, 60 lbs-inlet carton

  Assembled Space Requirement

  30" round * 64" high

  Power Requirement

  Single phase with earth ground

  110v/60hz or 220v/50hz

  Motor Specifications

  110v/60hz-part #TE-115923, 2-stage vacuum 1.0 HP,

  double ball bearing thru-flow discharge

  220v/50hz-part #TE-116111, 2-stage vacuum 1.0 HP,

  ball/sleeve bearing thru-flow discharge

  Altitude Application

  to 10,000 feet above sea level

  Ambient Temperature Application

  -30 to +50¡É

  Relative Humidity Application

  0-100%

  Sample Delay Start Range

  0-7 days

  Start up amps

  20 amps

  Running amps

  7 amps