Harim Engineering, Inc.

 

  

      Home / About Harim / Product / User List / Site Link

Korean Home

 

 

 

 

 

  TRS (API MODEL 102A) Analyzer

 

*0-50 ppb to 0-20 ppm ranges, user-selectable

*Dual Ranges and Auto-Ranging

*Microprocessor-based with simple, multitasking,   menu-driven software

*Test and Diagnostic Functions - continuous   self-checking with alarms

*Flexible Data Acquisition System for a variety of logging   and diagnostic functions

*Analog Outputs plus Serial Communication with APICOM   Windows-based operating software

*Optional Internal Zero and Span Sources

*Temperature and Pressure Compensated

      Specifications

UNITS:

ppm, ppb, mg/m3, micro g/m3

ZERO NOISE:

<0.2 ppb (RMS)

SPAN NOISE:

<0.5% of reading (above 50 ppb)

LOWER DETECTABLE LIMIT (LDL)

0.4ppb

ZERO DRIFT:

<0.5 ppb/24 hours, <1 ppb/7 days

SPAN DRIFT:

<0.5% reading/24 hours , <1% of reading/7days

PUMP:

Internal Diaphragm Type

RISE AND FALL TIME:

<120 seconds to 95%

LINEARITY:

1% full scale

PRECISION:

0.5% of reading

SAMPLE FLOW RATE

700cc/min +10%

OPERATING TEMPERATURE RANGE:

5-40oC

DIMENSIONS (HxWxD):

7" (178mm) x 17" (432mm) x 23.5" (597mm) (Analyzer)

7" (178mm) x 17" (432mm) x 22" (559mm) (Converter)

WEIGHT:

45 lbs. (20.5 kg) (Analyzer)
18
lbs. (8 kg) (Converter)

POWER:

100V 50/60Hz, 220V 50/60Hz, 230V 50/60Hz, 240V 50Hz - 250 watts

Converter:

External Thermal Type (800oC)

OUTPUTS (STANDARD):

Analog: Bi-Polar, 0-1/5/10V/100mV, selectable -

Serial: RS-232 (standard DB-9 connector)

OUTPUTS (OPTIONAL):

0-20mA or 4-20mA isolated

RECORDER OFFSET:

+10%