Harim Engineering, Inc.

 

  

      Home / About Harim / Product / User List / Site Link

Korean Home

 

 

 

 

 

TE-1000 PUF(poly-Urethane Foam) Monitor in Ambient Air

        TE-1000                        PUF   for Pesticide particulate/Vapor Air Sampler: Complete
                                                       with outdoor anodized aluminum PUF shelter, aluminum
                                                       blower motor assembly, 4" Particulate/Vapor sampling
                                                       module, flow venturi calibration valve system, 0-100"
                                                       magnehelic, gage, motor voltage control/elapsed time
                                                       indicator, 10" exhaust hose, and 7-day mechanical timer
                                                       (110v/60hz or 220v/50hz)

SPECIFICATIONS

  Inlet collection efficiency

  Meets EPA Methods TO4A, TO9A, TO13A

  Flow rate

  280Liters per min.(10 CFM)

  Filter media

  4" Round Quartz Filters(TE-QMA4)

  Flow control

  Motor Voltage Control/Elapsed Time Indicator  (TE-5010)

  Motor blower

  2-stage vacuum Moter 110v/60hz (TE-116336)

                                      220v/50hz (TE-116125)

  Flow Indicator

  0 to 100" Magnehelic Gage(TE-1005)

  Mechanical programmable timer

  7 Day Mechanical Timer (TE-5007)

  Elapsed time indicator device

  Elapsed time indicator(TE-5010)

  Shipping Information

 

  Size

  22.5"*20"*45.5" sampler carton

  14" * 14"* 19"lid holder carton

  Weight

  76 Ibs. - sampler carton,  8 Ibs. - lid carton

  Assembled Space Requirement

  24"square * 53" high

  Power Requirement

  Single phase with earth ground

  110v/60hz or 220v/50hz

  Motor Specifications

  110v/60hz  - Part #TE-116336, vacuum
  Double Ball Bearing.Bypass: Peripheral Discharge

  220v/50hz  - Part #TE-116125, vacuum
  Double Ball Bearing.Bypass: Peripheral Discharge

  Altitude Application

  to 10,000 feet above sea level

  Ambient Temperature Application

  -30 to +50¡É

  Relative Humidity Application

  0-100%

  Sample Delay Start Range

  0-7 days

  Start up amps

  12 amps

  Running amps

  8 amps