Harim Engineering, Inc.

 

  

      Home / About Harim / Product / User List / Site Link

Korean Home

 

 

 

 

 

 Mass Flow Calibrator
API 700

Photometric Ozone Calibrator
 API 401

 Repeatability of  Flow  Control

±0.2%full scale

Full scale Ranges

100ppb to 10ppm

 Linearity of Flow

 Measurements

±0.5% full scale

LDL

0.0006ppm

 Flow range of  Cylinder  Gases

0 to 100 cc/min

Zero Drift

<0.1ppb/7days

 Calibration Gas

 Input Port

4

Temperature

range

5 - 40¡É

 Flow Measurement

 Accuracy

±1% full scale

Precision

1.0ppb

 Approvals

CE 

Approvals

CE

 Application

Ambient

Monitoring 

Application

Low level test chamber

US EPA approved

Precision Calibrator
API 702

Zero Air System
API 701

Calibration gas ports

2

Output

20SLPM@30psi

Diluent flow

1 to 3 SLPM

Dew point

<-20¡É up to 15 LPM

<-10¡É up to 29 LPM

Flow stability

1%/7day

pump

Internal, oil-less

diaphragm, long life

Temperature range

5 - 40¡É

Weight

55lbs(25kg)

Calibration gas flow

10 to 50 cc/min

Approvals

CE 

Approvals

CE 

Application

Ambient

Monitoring 

Application

Ambient

Monitoring